product-detail-img

MPS2SENS nitrogen monoxide

PS2x4C2-A-U-S1.5-Kr-L1/D2 | 4594.63-H.62

2 channel premium pyro detector to measure nitrogen monoxide

  • Highly sensitive and long-term stable
  • High detectivity: up to 4,0*108 cm√Hz/W
  • High sensitivity: up to 200000 V/W
  • Measuring gas: NO (CWL: 5300 ± 30 nm/ HBW: 200 ± 20 nm)
  • Applications: e.g. aerospace, industrial, production and process engineering, environmental technology, process monitoring
from
77,70 €
Available
Save for later

Bulk prices

starting from 10 165,11 €
starting from 25 99,56 €
starting from 50 77,70 €

Product available with an extended delivery time of. Further informations ans help with procurement via our Micro Hybrid Sales Team

You are currently ordering as a guest. To enjoy all the benefits of our Plattform such as higher quantities, customized prices, inventory, tracking (order status), order overview or change of repository etc. please log in to My Micro Hybrid


Description

The 2 channel premium pyrodetector is specially designed for the analysis of nitric oxide (NO). The optical filter with a CWL of 5300 ± 30 nm and a HBW of ± 20 nm as well as an additional reference channel (CWL: 3910 ± 28, HBW: 70 ± 10 nm) optimize the measurement results of the dual IR detector. Its main application lies in the measurement of process gases or exhaust gases in industrial production and process engineering. Storage and operation can take place between -20 and 85 °C.

The particularly high detectivity of up to 4.0*108 cm√Hz/W as well as the very high sensitivity up to 200000 V/W provide an extraordinary signal quality in the sector of pyroelectric detectors. Two sensor chips with a thin film CPM membrane form the core of the dual detector MPS2SENS nitrogen monoxide. Thanks to the Ceramic Pyroelectrical Material (CPM) and the unique product design the detector features a high read-out rate up to 100 Hz as well as a short response time.

Please refer to the data sheet for detailed technical information. Customization services can be offered for this product as well as agreements on graduated prices.

Technical Data

Active area 2 x (0.7 x 0.7) mm²
Aperture 2 x (1.5 x 1.5) mm²
Mode
Current
Thermal time constant 16 ms
Electrical time constant 2 ms
AC sensitivity 200,000 V/W
Noise density voltage 50 µV/√Hz
Typical noise equivalent power (NEP)
typ. 4,0 x 10-10
Specific detectivity
Microphonic voltage typ. < 1000 μV/g
Optical filters
NO
Central wavelength: 5,300 nm
Half window width: 200 nm
Transmission: > 70 %
Filter code: L1
Blocking minimum: 400 nm
Blocking maximum: 11,000 nm
Target gas: nitrogen monoxide (NO)
Reference 3910 nm
Central wavelength: 3,910 nm
Half window width: 70 nm
Transmission: > 70 %
Filter code: D2
Blocking minimum: 400 nm
Blocking maximum: 11,000 nm
Operating voltage Vdd +5 ± 10 % V DC
Operating voltage VOffset 2.5 V DC
Maximum operating current 10 mA
Minimum operating temperature -25 °C
Maximum operating temperature 85 °C
Socket TO39