product-detail-img

MTS4SENS200 NOX

TS4x200B-A-S1.5-1-Kr-I1/L1/H1/D5 | 4594.40-2.A2

4 channel premium IR thermopile detector for measurement of nitrogen compounds

  • High sensitivity and long-term stability
  • High detectivity: up to 6,27*108 cm√Hz/W when filled with krypton
  • High sensitivity: up to 171 V/W
  • Target gases: H2O (filter, CWL: 6580 ± 40 nm / HBW: 200 ± 35 nm), NO (filter, CWL: 5300 ± 30 nm / HBW: 200 ± 20 nm), SO2 (filter, CWL: 7300 ± 50 nm / HBW: 180 ± 20 nm)
  • Applications: e.g. industry / raw materials and resources, environmental technology
from
81,11 €
Available
Save for later

Bulk prices

starting from 10 126,75 €
starting from 25 92,52 €
starting from 50 81,11 €

You are currently ordering as a guest. To enjoy all the benefits of our Plattform such as higher quantities, customized prices, inventory, tracking (order status), order overview or change of repository etc. please log in to My Micro Hybrid


Description

This 4 channel detector is tuned to the wavelength of nitrogen oxides with 4 optical filters. The optical filters determine the infrared detection of the following gases: water vapor (H2O) (CWL: 6580 ± 40, HBW: 200 ± 35 nm), nitric oxide (NO) (CWL: 5300 ± 30, HBW: 200 ± 20 nm), sulfur dioxide (SO2) (CWL: 7300 ± 50, HBW: 180 ± 20 nm).

The four-channel IR thermopile for NDIR gas analysis has 4 sensor chips in the core, each with 200 BiSb/Sb thermopairs. The signal strength of the thermopile detector is comparable to that of a pyrodetector: high detectivity up to 6.27*108 cm√Hz/W and high sensitivity up to 171 V/W with the fill gas krypton. The interference absorber, an integrated thermistor for reference temperature determination and the reference channel (CWL: 5060 ± 25, HBW: 100 ± 15 nm) are further important product features for long-term use in demanding applications.

The MTS4SENS200 NOX provides reliable measurement values in ambient temperatures between -20 and 85 °C, e.g. for industrial gas measurement or process analysis in a wide range of industrial and manufacturing sectors, such as chemicals, textiles, environmental technology, recycling or exhaust-gas technology.

For further graduated prices and possible individual, technical adaptations, our sales team will be happy to advise you.

Technical Data

Active area 4 x (1,2 x 1,2) mm²
Aperture 4 x (1,5 x 1,5) mm²
Number of thermal pairs 200
Time constant 51 ms
DC output voltage 9.35 mV
DC sensitivity 171 V/W
Noise voltage 33 nV/√Hz
Noise equivalent power (NEP) 0.19 nW/√Hz
Specific detectivity
Resistance of thermal coil 65 kΩ
Thermistor
Detector filter
H2O
Central wavelength: 6,580 nm
Half window width: 200 nm
Transmission: > 70 %
Filter code: I1
Blocking minimum: 100 nm
Blocking maximum: 14,000 nm
Target gas: water vapour (H2O)
NO
Central wavelength: 5,300 nm
Half window width: 200 nm
Transmission: > 70 %
Filter code: L1
Blocking minimum: 400 nm
Blocking maximum: 11,000 nm
Target gas: nitrogen monoxide (NO)
SO2
Central wavelength: 7,300 nm
Half window width: 180 nm
Transmission: > 80 %
Filter code: H1
Blocking minimum: 400 nm
Blocking maximum: 11,000 nm
Target gas: sulfur dioxide (SO2)
Reference 5060 nm
Central wavelength: 5,060 nm
Half window width: 100 nm
Transmission: > 70 %
Filter code: D5
Blocking minimum: 400 nm
Blocking maximum: 300,000 nm
Minimum operating temperature -20 °C
Maximum operating temperature 85 °C